金属清洗剂中氯离子检测方案(自动电位滴定)

收藏
检测样品: 洗涤剂
检测项目: 含量分析
浏览次数: 151
发布时间: 2022-07-19
关联设备: 2种 查看全部
获取电话
留言咨询
方案下载
方案详情 产品配置单
金属清洗剂中氯离子的测定 应用资料 用电位滴定法测定金属清洗剂中的氯离子浓度。首先,向样品中加入水和硝酸溶液,然后用0.1mol/L硝酸银溶液滴定至终点。终点是滴定曲线上的最大拐点。根据硝酸银的滴定体积计算氯离子的浓度。

方案详情

KGM KYOTO ELECTRONICSMANUFACTURING CO.,LTD.TIQ-97057enL-K-M KYOTO ELECTRONICSMANUFACTURING CO.,LTD. Application MemoChlorine Ion in Metal Cleanser 1.Overview Chlorine ion concentration in metal cleanser is measured by potentiometric titration.First, thesample is added with water and nitric acid solution, and then titrated with 0.1mol/L silvernitrate solution up to the endpoint. The endpoint is the maximum inflexion on the titrationcurve. The concentration of chlorine ion is calculated from the titration volume of silvernitrate. 2. Apparatus Main unit Automatic potentiometric titrator (preamplifier STD) Electrode Silver electrode Mercury sulfate reference electrode 3. Reagents Titrant 0.1mol/L silver nitrate solutionAdditive Nitric acid solution 4. Example Sample Titer Chlorine ion (g) (mL) (ppm) 1 2 3 5.1315 1.2638 874.30 5.1138 1.2610 875.39 5.1206 1.2589 872.77 AverageSD RSD(%) 872.15 1.32 0.151 -Titration curve- Please feel free to contact us for any further information. Kyoto Electronics Manufacturing Co., Ltd. Overseas Sales & Marketing Sect. 京都电子工业株式会社-可睦电子(上海)商贸有限公司 上海市徐汇区宜山路333号1201室电话:021-54488867 电邮: kemu-kem@163.com 网址: http://www.kem-china.com 金属清洗剂中氯离子的测定 应用资料(英文版)用电位滴定法测定金属清洗剂中的氯离子浓度。首先,向样品中加入水和硝酸溶液,然后用0.1mol/L硝酸银溶液滴定至终点。终点是滴定曲线上的最大拐点。根据硝酸银的滴定体积计算氯离子的浓度。
确定

还剩1页未读,是否继续阅读?

不看了,直接下载
继续免费阅读全文

该文件无法预览

请直接下载查看

可睦电子(上海)商贸有限公司-日本京都电子(KEM)为您提供《金属清洗剂中氯离子检测方案(自动电位滴定)》,该方案主要用于洗涤剂中含量分析检测,参考标准--,《金属清洗剂中氯离子检测方案(自动电位滴定)》用到的仪器有AT-710M四通道旗舰型自动电位滴定仪、AT-710S豪华型自动电位滴定仪