核心参数
产品种类: 自动旋光仪
温控功能: 可控温
数显功能: 有数显
光源类型: 钠光灯
产地类别: 进口
测量范围: ±90°
测量精度: 0.0001°
测量准确度: ±0.002°
旋光仪 型号:P-2000
概要
P-2000设计为可定制的旋光仪,具有各种选项,适用于各种应用。 系统也可以根据应用需求的变化进行现场升级。 滤光片和光源的选项提供了从紫外-可见到近红外的广泛的分析波长。
规格
型号名称 | 2100 | 2200 | 2300 |
光源 (选项) | 卤素灯 (钠灯?水银灯) | 钠灯 (水银灯?卤素灯) | 钠灯?水银灯 (卤素灯) |
偏光片 | Glan–Taylor | Glan–Taylor | |
滤光片波长 | 589nm(选项波长对应) | 589, 578, 546, 436, 365nm (选项波长对应) | |
检测器 | 光电子倍增管 | ||
最大使用样品池长度 | 100mm | ||
测量角度范围 | ±90° | ||
测量准确性 | ±0.002°(~1°)、±0.2%(1°~) | ||
测量再现性 | ±0.002° | ||
最小表示角度 | 0.0001° | ||
响应速度 | 6°/sec | ||
测量方式 | 対称角震动方式光学零位法 | ||
识别功能 | 光源、滤光片、附件 | ||
温度测量点 | 样品池支架/样品池内 | ||
温度测量范围 | 0~99.9℃(15~40℃的范围温度正确性±0.1℃) | ||
尺寸 | 635(W)×364(D)×249(H)mm | ||
重量 | 25kg | 28kg | 30kg |
日本分光计器旋光仪P-2000的工作原理介绍
旋光仪P-2000的使用方法?
日本分光计器P-2000多少钱一台?
旋光仪P-2000可以检测什么?
旋光仪P-2000使用的注意事项?
日本分光计器P-2000的说明书有吗?
日本分光计器旋光仪P-2000的操作规程有吗?
日本分光计器旋光仪P-2000报价含票含运吗?
日本分光计器P-2000有现货吗?
更多
更多
更多
用钠灯和卤素灯测定吡柔比星的旋光
An emission line of light source such as a sodium lamp or a mercury lamp is usually used to measure the optical rotation of pharmaceutical products. However, the Japanese Pharmacopeia and the European Pharmacopeia only mention the use of a sodium lamp as the light source. A polarimeter using a halogen lamp and a band pass filter (BPF) can also be used, as mentioned in the United States Pharmacopeia. However, the measurement error from the transmission property, that is the difference between the center wavelength of the BPF and the wavelength of emission line, cannot be avoided. Herein, the optical rotation of pirarubicin was measured using a sodium lamp and a halogen lamp, and the measurement error between the results was evaluated.
制药/生物制药
2019/12/22
用汞灯365 nm测量氢溴酸右美沙芬的旋光
Optical rotation of Dextromethorphan Hydrobromide in accordance with the U.S. Pharmacopeia, is requested to be measured at 325 nm, however there is no emission line of 325 nm available in the polarimeter using Na lamp or Hg lamp as light source. Therefore, for the measurement of optical rotation of Dextromethorphan Hydrobromide, a polarimeter with halogen lamp as a light source and interference filter (325 nm) has been usually used, but as shown in Fig. 1, the optical rotatory dispersion (ORD) spectrum, there is a very steep slope in the short wavelength range near 325 nm, where even a very small error in the wavelength of interference filter may cause a very large error in the optical rotation measurement, which will makes the accurate measurements very difficult. On the other hand, if the measurement by using 365nm is allowed, since no wavelength error is expected, the measurement of optical rotation can be implemented with much higher accuracy. In this experiment, Dextromethorphan Hydrobromide was measured using polarimeter with Hg lamp (365 nm), polarimeter with Halogen lamp and interference filter (325 nm) and polarimeter with ORD attachment.
制药/生物制药
2019/12/22