Gaertner LSE-2A2W 进口椭偏仪

LSE-2A2W Ellipsometer specifications

Alignment:

Tilt and table height using computer alignment screen.

Incidence Angles:

fixed at  60° for 544nm laser and 70°  for 633nm   laser

Method of Measurement:

Advanced StokesMeter determines the complete measuring beam   polarization using no moving parts and no modulators, only 4 stationary   silicon detectors so measurements are stable and exact.

Measurement Time:

Partically instantanious

Measuring Laser:

544 & 633 nm HeNe Gas Lasers with 1 mm measuring beam diameter

Alignment Laser:

670nm Laser Diode

Sample (Wafer) Size:

300    mm wide X unlimited length

Stage:

Hand positioning with tilt and table height adjustment.

Software:

Windows LGEMP 4   layer absorbing included.

Required Computer:

(not included) Windows  PC or laptop with USB port

Film Thickness Range:

0 - 60,000 Angstroms on substrate or on 1, 2, 3, or 4 known   sublayers.

Precision, Repeatability:

Sub-Angstrom over most of the measurement range.

Refractive Index:

± .001 over most of the measurement range.

Power:

Supplied by computer USB bus and laser power supplies.

Dimensions:

Height:  16.5   Width: 27.5   Depth: 16   inches
  Net Weight: 45 lbs.    Shipping Weight: 65 lbs.

CDRH Compliance:

All laser ellipsometers supplied by Gaertner comply with CDRH   requirements 21CFR 1040 for a Class II laser product emitting less than 1 mW   or Class IIIb less than 5 mW of low power radiation. As with any bright   source such as the sun or arc lamp, the operator should not stare directly   into the laser beam or into its reflection from highly reflecting surfaces.

CE Compliance:

Comply with European safety directives and carry the CE mark.

 

 


Model LSE Ellipsometer

 
Manufactured by Gaertner Scientific, USA
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Stokes Ellipsometer LSE Description
The Stokes Ellipsometer LSE is a continuation of an entirely new line of ellipsometers based on the advanced StokesMeter technology - winner of Photonics Spectra and R&D 100 best new products awards.
The LSE model ellipsometer includes data acquisition and analysis software.  An optional Windows based computer with PCI interface is available.
This ellipsometer uses patented StokesMeter technology with no moving parts and no modulators to quickly and accurately determine the complete polarization state of the 6328Å laser measuring beam at a 70° incidence angle. The space-saving design features a small footprint yet it can accommodate large samples up to 300mm wide. The sample stage can be easily moved by hand to measure any point on the sample surface. The sample table includes a manual tilt and table height adjustment which is set using an alignment screen on the computer.
The data acquisition and analysis software, LGEMP, can measure the top layer film thickness and film refractive index on a substrate or on 1, 2, or 3 known bottom layers. The films can be transparent or absorbing. This durable integration of hardware and software is fast and easy to use.
Extremely precise, stable and low cost the model LSE Stokes Ellipsometer represents an excellent value in a basic ellipsometer.
Description of Stokesmeter Technology
This patented device uses no moving parts and no modulators to quickly and accurately determine the complete polarization state of the measuring beam.


The diagram above shows the StokesMeter photopolarimeter for the simultaneous measurement of all four Stokes parameters of light. The light beam, the state of polarization of which is to be determined, strikes, at oblique angles of incidence, three photodetector surfaces in succession, each of which is partially spectrally reflecting and each of which generates an electrical signal proportional to the fraction of the radiation it absorbs. A fourth photodetector is substantially totally light absorptive and detects the remainder of the light. The four outputs thus developed form a 4x1 signal vector I which is linearly related, I=AS, to the input Stokes vector S. Consequently, S is obtained by S=A(-1)I. The 4x4 instrument matrix A must be nonsingular, which requires that the planes of incidence for the first three detector surfaces are all different. For a given arrangement of four detectors, A can either be computed or determined by calibration.
Technological Advantage
The clean, compact StokesMeter replaces a typical rotating analyzer assembly consisting of a drum, prism, encoders, switches, motor and detector and their associated electronics. In addition, the waveplate mechanism on the polarizer arm, is eliminated. This results in a fast, precise, stable no moving parts ellipsometer.
An outstanding feature is the Stokes Ellipsometer's compensation for small changes in angular beam deviation caused by sample out-of-flatness. This permits fast, uninterrupted measurement over the entire wafer surface without the need to pause to correct for focus and tilt. When scanning similar samples, tilt-free, focus-free operation is the obvious benefit.
 
LSE Ellipsometer specifications

Alignment: Tilt and table height on computer alignment screen.
Incidence Angle: 70°
Method of Measurement: Patented StokesMeter determines the complete beam polarization using no moving parts and no modulators, only 4 stationary silicon detectors so measurements are exact and stable.
Measurement Time: Practically instantaneous.
Light Source: 6328Å HeNe Laser
Beam Diameter: 1 mm
Sample (Wafer) Size: 300 mm wide X unlimited length
Stages: Hand positioning with tilt and table height adjustment.
Software: Windows 2000/XP LGEMP 4 layer absorbing.
Computer: Windows 2000/XP PC with available PCI slot (option) 
Film Thickness Range: 0 - 60,000 Angstroms on substrate or on 1, 2, 3, or 4 known sub-layers.
Precision & Repeatability: Sub-Angstrom over most of the measurement range.
Refractive Index: ± .001 over most of the measurement range.
Power: Supplied by computer interface card and laser power supply.
Dimensions: Height:  16.5   Width: 27.5   Depth: 16 inches
Net Weight: 40 lbs.  Shipping Weight: 65 lbs.
CDRH Compliance: All laser ellipsometers supplied by Gaertner comply with CDRH requirements 21CFR 1040 for a Class II laser product emitting less than 1 milliwatt of low power radiation. As with any bright source such as the sun or arc lamp, the operator should not stare directly into the laser beam or into its reflection from highly reflecting surfaces.
CE Compliance: S series Ellipsometers comply with European safety directives and carry the CE mark.

 
 中国销售服务商
美国赛伦科技上海办事处
吴惟雨




售后服务
保修期: 根据合约
是否可延长保修期:
现场技术咨询:
免费培训: 根据合约
免费仪器保养: 根据合约
保内维修承诺: 根据合约
报修承诺: 根据合约
典型用户

用户单位

采购时间

采购数量

无锡华润华泾

2011/06/07

2

TPK(厦门)

2011/05/15

12

问商家

Gaertner椭偏仪Model LSE-2A2W, LSE的工作原理介绍

椭偏仪Model LSE-2A2W, LSE的使用方法?

GaertnerModel LSE-2A2W, LSE多少钱一台?

椭偏仪Model LSE-2A2W, LSE可以检测什么?

椭偏仪Model LSE-2A2W, LSE使用的注意事项?

GaertnerModel LSE-2A2W, LSE的说明书有吗?

Gaertner椭偏仪Model LSE-2A2W, LSE的操作规程有吗?

Gaertner椭偏仪Model LSE-2A2W, LSE报价含票含运吗?

GaertnerModel LSE-2A2W, LSE有现货吗?

工商信息

企业名称

赛伦科技(北京)有限责任公司

企业信息已认证

企业类型

有限责任公司(外商合资)

信用代码

110000410283047

成立日期

2006-03-06

注册资本

美元8万元

经营范围

许可经营项目:无一般经营项目:电子技术、新能源技术的研发及技术咨询;销售自行研发的产品;批发电子设备、半导体设备、集成电路加工设备、集成电路检测设备及以上产品的零备件、科学仪器及仪表、家用电器,提供上述产品的技术服务.(未取得行政许可的项目除外)(知识产权出资0.8万美元。)理富,才清华

联系我们
赛伦科技为您提供美国Gaertner椭偏仪/膜厚测量仪/EllipsometerModel LSE-2A2W, LSE,GaertnerModel LSE-2A2W, LSE产地为美国,属于进口椭偏仪,除了美国Gaertner椭偏仪/膜厚测量仪/Ellipsometer的参数、价格、型号、原理等信息外,还可为您提供进口四探针电阻率测试仪(4PP)/方块电阻测试仪,赛伦科技客服电话400-860-5168转2024,售前、售后均可联系。

赛伦科技

查看电话

沟通底价

提交后,商家将派代表为您专人服务

获取验证码

{{maxedution}}s后重新发送

获取多家报价,选型效率提升30%
提交留言
点击提交代表您同意 《用户服务协议》 《隐私政策》 且同意关注厂商展位
联系方式:

公司名称: 赛伦科技(上海)办事处

公司地址: 上海市黄浦区陆家浜路1378号万事利大厦601室 联系人: 吴13817915874 邮编: 200011 联系电话: 400-860-5168转2024

友情链接:

仪器信息网APP

展位手机站