Rheoline MD
Prescott Instruments' Rheo-Line has been developed to meet the needs of a modern testing laboratory, providing fast and accurate data making it ideally suited to both the production environment and for use in the development of elastomeric compounds. Samples can be tested prior to the curing process and their properties measured allowing accurate prediction of product or material durability.
At the heart of the Rheo-Line system is a rotor- less Biconical Die assembly with heat being applied to upper and lower dies producing precise accurate temperature control. The lower die is oscillated at 1.66 Hz at an amplitude of 0.5°, 1.0° or 3.0° while the top die remains fixed and accurately measures the reaction torque.
Rheoline MD
Technical Specifications:
Standard: Conforms to ISO No. 6502 / ASTM D5289
Electrical: Single Phase, 220 / 240v 50Hz, or 110v 60Hz
Pneumatics: Filtered Air 4.2 Kg/cm² ( 6OPSI )
Cavity Configuration: Sealed or non sealed die system
Oscillation Frequency: 1.67 Hz
Oscillation Amplitude: 0.5°, 1.0°, 3.0°. (0.5° fitted as standard)
Temperature Control: Digital PID
Temperature Range: 30° to 250°C
Torque Units: In/Lbs or dNm
Temperature Units: Celsius or Fahrenheit
Pressure Units: Lb/In² or Kg/cm² (optional)
Time Units: Min/Seconds or Minutes/Decimal
Real Time Graphical Output: Elastic Modulus, Viscous Modulus, Tan Delta,Cure Rate,Pressure in Die cavity (optional)
PC Specification: Pentium processor, Fully Networkable
Data Format: All data stored in accessible format allowing easy export to all other packages.
Optional Extras Available: Programmable temperature zones
Additional oscillation eccentric
Pressure transducer
Auto loading module (see Automated Rheo-Line Moving Dir Rheometer)
MDR Sample Cutter
Rheoline MD
Prescott Instruments' Rheo-Line has been developed to meet the needs of a modern testing laboratory, providing fast and accurate data making it ideally suited to both the production environment and for use in the development of elastomeric compounds. Samples can be tested prior to the curing process and their properties measured allowing accurate prediction of product or material durability.
At the heart of the Rheo-Line system is a rotor- less Biconical Die assembly with heat being applied to upper and lower dies producing precise accurate temperature control. The lower die is oscillated at 1.66 Hz at an amplitude of 0.5°, 1.0° or 3.0° while the top die remains fixed and accurately measures the reaction torque.