Restek 测苯甲酸的方法
2010-08-31 10:38
浏览:472次
分享:
资料摘要:
US EPA Method 8270D by GC/MS
Rtx®-XLB
Rtx®-XLB, 20m, 0.18mm ID, 0.18μm (cat.# 42802)
Sample: US EPA Method 8270D mix:
8270 MegaMix™ (cat.# 31850),
benzoic acid (cat.# 31415),
benzidine (cat.# 31441),
2,4-dinitrophenol (cat.# 31291),
Acid Surrogate Mix (4/89 SOW)
(cat.# 31063),
B/N Surrogate Mix (4/89 SOW)
(cat.# 31062)
Inj.: 0.5μL, 5ppm each analyte
(2.5ng on column) (2.5ppm/1.25ng
on column for 3-methylphenol and
4-methylphenol)
2mm splitless cyclo double
gooseneck injector liner
(cat.# 20907);
splitless hold time 0.15 min.;
pressure pulse: 0.20 min. @30psi
GC: Agilent 6890
Inj. temp.: 270°C
Carrier gas: helium
Flow rate: 1.2mL/min., constant flow
Oven temp.: 40°C (hold 0.5 min.) to 90°C @
14°C/min. (no hold) to 330°C @
22°C/min. (hold 1 min.)
Det.: Agilent 5973 GC/MS
Transfer line
temp.: 280°C
Scan range: 35–550 amu
Solvent delay: 1 min.
Tune: DFTPP
Ionization: EI
相关资料
Raptor C18 柱分析全氟烷酸化合物,可将你EPA 方法537 的分析时间缩短将近半个小时!
且在10分钟完整周期时间内具有极好的分离效果;可为很多应用,基质和化合物提供了极好的数据。
色谱柱的安装,色谱柱的储存,色谱柱的再生,MS 型色谱柱的老化,PLOT 色谱柱的安装和老化,故障排解,色谱柱退化的原因.
气相色谱柱(以下简称色谱柱或柱子),是气相色谱(质谱)分析设备的心脏部件。选择一根适用的色谱柱不仅仅 能够对样品进行有效的分离,更是色谱分析结果准确性和可靠性的保证。如果使用了不合适的色谱柱产品,得到的是不 完全的分离结果和不准确的定性定量结果。
大约有10,000 种化合物可以用约400 种不同种类气相色谱柱进行分离。对于这样数量繁多的应用项目,色谱柱生产制 造厂家需要帮助用户选择到实际需要的产品。GS-TEK 通过多年的色谱柱制造和应用经验,可以快速地帮助您选择到合适 的色谱柱产品。
阿莫西林在Ultra IBD柱完美分离,金霉素、强力霉素、四环素在Ultra C18柱上的完美分离,磺胺二甲基嘧啶 硫抗生素在Raptor Biphenyl柱上完美分离