恩德斯豪斯非玻璃数字式pH电极Memosens CPS97E
产品参数测量原理ISFET电极应用Unbreakable pH sensor for use in highly polluting media:-Chemical industry-Pulp and paper industry-Mineral and metal industry特点Digital pH electrodes for with open aperture for clogging media e.g. suspended solides, dispersion, precipitation reactionNeeds min. conductivity of 500μS/cmReferenceStabilized gel reference system测量范围pH 0 to 14测量原理测量原理设计设计材质Sensor shaft: PEEKSealings: FFKMMetal lead: Ag/AgClO-ring: FKMProcess coupling: PPS fiber-glass reinforcedNameplate: Ceramic metal oxide外形尺寸Diameter: 12 mm (0.47 inch)Shaft lengths: 120, 225, 360 and 425 mm(4.72, 8.86, 14.2 and 16.7 inch)过程温度-15 to 110 °C (5 to 230 °F)过程压力Max. 11 bar abs. at 100 °C(Max. 159.5 psi at 212 °F)温度传感器Pt1000防爆认证With the following approvals for use in potentially explosive areas of Zone 0, Zone 1 and Zone 2: ATEX, IECEx, CSA C / US, NEPSI, JPN Ex, INMETRO,UKCA and Korea Ex连接Inductive, contactless connection head with Memosens 2.0 technology防护等级IP 68应用领域Memosens CPS97E pH电极,无玻璃破裂风险,特别适合高纤维含量介质和固体颗粒悬浮液测量,比如分散液、沉淀液和乳液。非玻璃电极杆,无破裂风险,带开孔隔膜,能够在严苛工况下可靠测量。CPS97E采用Memosens 2.0数字技术,支持扩展数据存储,保存标定参数和过程参数,支持预维护。电极防潮防水,允许实验室传感器标定,提高过程完整性,缩短生产中断时间。化工行业• 污染介质和固体颗粒悬浮液- 分散液- 沉淀液- 乳液纸浆和造纸行业优势Memosens 2.0支持扩展储存单元,保存标定参数和过程参数,能够更好地进行趋势识别,是实施预维护和推进IIoT技术的完美基础。无玻璃破裂风险:PEEK电极杆,耐受严苛工况应用,确保产品安全开孔隔膜满足重度污染介质测量的要求,防止电极堵塞。参比凝胶化学性质稳定,保证测量结果长期稳定低维护工作量:ISFET技术不受高温或温度波动的影响,标定间隔时间长最高过程安全性:采用非接触式感应式信号传输方式,潮湿环境和腐蚀工况对测量没有影响。成功案例监测乳液的聚合反应通过反应结束检测和动力学控制实现工艺优化半间歇聚合工艺拉曼分析仪采用Kaiser拉曼技术,用于监测和控制半间歇聚合工艺