邻甲苯胺及其同分异构体的分离The Separation of O-Toluidine and Its isomers摘要:为了实现邻甲苯胺及其同分异构体的完全分离,首先,作者将邻甲苯胺及其同分异构体乙酰化,然后采用气相色谱-质谱联用方法,通过改进色谱条件,使其取得了完全分离。试验表明采用气-质联用方法可以实现邻甲苯胺及其同分异构体的完全分离。关键词:邻甲苯胺;分离;异构体;乙酰化Abstract: In order to achieve the complete separation of o-toluidine and its isomers, firstly, o-toluidine and its isomers were acetylation, and then the authors used gas chromatography-mass spectrometry methods, by improving the chromatographic conditions, so as to achieve their completely separation. The results showed that the gas-chromatography-mass spectrometry method can achieve the completely separation of o-toluidine and its isomers.Keywords: o-toluidine; separation; isomers; acetylation1 前言在禁用偶氮染料检测过程中,邻甲苯胺是经常被检出者之一。邻甲苯胺有两个同分异构体,分别是间甲苯胺和对甲苯胺。其中,邻甲苯胺属于禁用芳香胺,而间甲苯胺和对甲苯胺则不属于禁用芳香胺。三个化合物的沸点非常接近,分别为:邻甲苯胺(200±2℃)、间甲苯胺(203±3℃)、对甲苯胺(201±1℃)。而且三者的质谱图很相似,只是分子极性上略有差别,故造成在用气相色谱-质谱联用方法检测到邻甲苯胺时经常出现假阳性结果。如不采用薄层色谱、液相色谱等其它分析手段共同鉴别,是无法判断是否假阳性结果的。但是要排除假阳性结果,就需要重新寻找条件,而且更换仪器费时费力,且多数实验室不一定同时具备这些设备,这就给我们的日常检测工作带来了很多的麻烦和不便。本文作者结合自身知识和经验,对常见的邻甲苯胺及其异构体分离问题进行了研究。通过查阅相关资料,作者发现邻甲基乙酰苯胺(296℃)、间甲基乙酰苯胺(303℃)和对甲基乙酰苯胺(307℃)三者的沸点相差较大,所以作者首先通过试验将邻甲苯胺及其异构体分别乙酰化,然后通过改进色谱条件,使以上化合物达到了完全分离,提高了检测效率,减少了检测过程中的假阳性检出。 2 试验2.1 仪器与试剂气相色谱-质谱联用仪(GC-MS):Agilent 7890A/5975C,美国Agilent公司毛细管柱:DB-17MS柱(30m×0.25mm×0.25μm)叔丁基甲醚、乙酸酐、无水碳酸钠和无水碳酸钾均为分析纯 旋转蒸发仪 上海亚荣生化仪器厂邻甲苯胺及其同分异构体均为德国Dr.Ehrenstorfer公司。2.2 试剂配制标准品溶液:用叔丁基甲醚为溶剂,分别称取适量邻甲苯胺、间甲苯胺和对甲苯胺标准品配成合适浓度的单标溶液及混合溶液。碳酸钾溶液:0.1mol.L-1水溶液,取13.8g[font
最近检测一些带印花的床上用品时发现,居然检出了5-硝基-邻甲苯胺,还有2,4-二氨基甲苯,有的十几ppm,有的都是好几十ppm,把阳性样品送到其他大型权威检测机构测试却只是2,4-二氨基甲苯不合格,而且比我检出的大,5-硝基-邻甲苯胺却是未检出!有没有同行出现过这种情况?求各位帮忙指点一下!1.标准也说了5-硝基-邻甲苯胺会分解成2,4-二氨基甲苯,按理说5-硝基-邻甲苯胺应该分解完的,不应该检出。会不会是保险粉的问题而导致反应不完全,从而5-硝基-邻甲苯胺没有完全分解?但是我做能力验证数据都很好,实验过程没有变化。(我的混标里面有加5-硝基-邻甲苯胺,所以能测出)2.我这些报告应该怎样出呢?这种情况已经遇到好几次了,都是印花部位!
偶用邻联甲苯胺比色法测水中余氯的含量,可是因为比色皿和比色管中经常粘有未冲洗干净的杂质,导致时间一长,测试结果就不准确,所以我想咨询下邻联甲苯胺用什么试剂清洗效果比较好呢?