5-1. Trace Nitrogen Analyzer
Model : PES-1001
Component to be measured : Trace Nitrogen in Argon gas
Measuring method : Emission spectrometric method
Measuring range : 0~5, 0~10 ppm N2(Manual select)
Repeatability : ±2.5% of full scale (By sequential measuring of standard gas with three times)
Linearity : ±2.0% of full scale with each range
Response time : Within 1 minute (90% response from analyzer inlet)
Sample gas flow rate : 1.0~2.0 L/min.(usually 1.0 L/min.)
Output signal : ★ DC 4~20mA (isolated linear signal)
★ DC 0~1V (non-isolated linear signal)
Display : Digital direct reading
Power source : AC 100V±10V, 50 Hz±1%, 250VA
Ambient temperature : 10~35℃