简介:2.1 Prior to analysis, samples which require total ("acid-leachable") values must be
digested using appropriate sample preparation methods (such as Methods 3005 - 3052).
2.2 Method 6020 describes the multi-elemental de简介:2.1 Prior to analysis, samples which require total ("acid-leachable") values must be
digested using appropriate sample preparation methods (such as Methods 3005 - 3052).
2.2 Method 6020 describes the multi-elemental determination of analytes by ICP-MS in
environmental samples. The method measures ions produced by a radio-frequency inductively
coupled plasma. Analyte species originating in a liquid are nebulized and the resulting aerosol is
transported by argon gas into the plasma torch. The ions produced by high temperatures are
entrained in the plasma gas and introduced, by means of an interface, into a mass spectrometer.
The ions produced in the plasma are sorted according to their mass-to-charge ratios and quantified
with a channel electron multiplier. Interferences must be assessed and valid corrections applied
or the data flagged to indicate problems. Interference correction must include compensation for
background ions contributed by the plasma gas, reagents, and constituents of the sample matrix.详细>