简介:2.1 Aqueous Samples: An inert gas is bubbled through a portion of the aqueous sample at
ambient temperature, and the volatile components are efficiently transferred from the aqueous phase
to the vapor phase. The vapor 简介:2.1 Aqueous Samples: An inert gas is bubbled through a portion of the aqueous sample at
ambient temperature, and the volatile components are efficiently transferred from the aqueous phase
to the vapor phase. The vapor is swept through a sorbent column where the volatile components
are adsorbed. After purging is completed, the sorbent column is heated and backflushed with inert
gas to desorb the components onto a gas chromatographic column.
2.2 High Concentration Extracts from Method 5035: An aliquot of the extract prepared in
Method 5035 is combined with organic free reagent water in the purging chamber. It is then
analyzed by purge-and-trap GC or GC/MS following the normal aqueous method.详细>