简介:2.1 This method provides the option of using either standard column chromatography
techniques or solid-phase extraction cartridges. Generally, the standard column chromatography
techniques use larger amounts of adsorbe简介:2.1 This method provides the option of using either standard column chromatography
techniques or solid-phase extraction cartridges. Generally, the standard column chromatography
techniques use larger amounts of adsorbent and, therefore, have a greater cleanup capacity.
2.2 In the standard column cleanup protocol, the column is packed with the required amount
of adsorbent, topped with a water adsorbent, and then loaded with the sample to be analyzed.
Elution of the analytes is accomplished with a suitable solvent(s) that leaves the interfering
compounds on the column. The eluate is then concentrated (if necessary).
2.3 The cartridge cleanup protocol uses solid-phase extraction cartridges packed with 1 g
or 2 g of silica gel (silicic acid) adsorbent. Each cartridge is solvent washed immediately prior to use.
Aliquots of sample extracts are loaded onto the cartridges, which are then eluted with suitable
solvent(s). A vacuum manifold is required to obtain reproducible results. The collected fractions
may be further concentrated prior to gas chromatographic analysis.
2.4 The appropriate gas chromatographic method is listed at the end of each technique.
Analysis may also be performed by gas chromatography/mass spectrometry (Method 8270).详细>