简介:This work describes a novel, simple, and inexpensive approach for an absorptive extraction of organic contaminants from aqueous samples
applying polysiloxane tubes. After the extraction, the tubes were desorbed with 200简介:This work describes a novel, simple, and inexpensive approach for an absorptive extraction of organic contaminants from aqueous samples
applying polysiloxane tubes. After the extraction, the tubes were desorbed with 200 Lof an organic solvent and 50 Lof the extract were transferred
via large volume injection (LVI) into a gas chromatography–mass spectrometry system (GC–MS). Fourteen organic compounds covering a broad
range in polarity with logKo/w-values between 2.6 (atrazine) and 7 (polychlorinated biphenyl 138, PCB 138) were chosen as model compounds.
The experimental parameters affecting the efficiency of the process such as desorption solvents, the addition of salt and pH-change, the extraction
and desorption time were carefully optimised. The recoveries were compared with theoretical values. The limits of detection (LODs) were in the
range of 0.5–5 ng/L and the precision was 7–15%. The method was successfully applied to the analysis of a contaminated river water sample.
详细>