简介:1.6.1. 元素分析范围从钾(K)到铀(U)(Elements can be analyzed from natrium (Na) to uranium (U) );
1.6.2. 元素含量分析范围为1 PPm到99.99%(Analysis contents of elements from 1 ppm to 99.9%);
1.6.3. 测量时间:60-200秒(Testin简介:1.6.1. 元素分析范围从钾(K)到铀(U)(Elements can be analyzed from natrium (Na) to uranium (U) );
1.6.2. 元素含量分析范围为1 PPm到99.99%(Analysis contents of elements from 1 ppm to 99.9%);
1.6.3. 测量时间:60-200秒(Testing time : 60-200 seconds);
1.6.4. ROHS指令规定的有害元素(限Cd/Pb/Cr/Hg/Br)检测其限度最高达1PPM;
( The limited detection hazardous elements(only aim at Cd/Pb/Cr/Hg/Br) up to 1PPM )
1.6.5. 多次测量重复性可达0.1%(总荧光强度);
The measurement repeatability on many occasions up to 0.1%。(all intensity of fluorescence)
1.6.6. 长期工作稳定性为0.1%;(总荧光强度)
Long-term working stability being 0.1%.(all intensity of fluorescence)
1.6.7. 能量分辨率为170±5电子伏特( Energy resolution :170±5EV ) ;
1.6.8. 温度适应范围为15℃至26℃ (Temperature of suitability from 15 °C to 25 °C );
1.6.9. 电源:交流220V±5V;(建议配置交流净化稳压电源。)
Power supply. AC 220 V ± 5 V;( AC purified stabilized voltage power supply is suggestion)
1.6.10.相互独立的基体效应校正模型( The correction model of relative independent matrix effect );
1.6.11.多变量非线性回归程序 ( Multivariate nonlineartly recycle program );
1.6.12.任意多个可选择的分析和识别模型 (Arbitrary optional analysis and identification model );
1.6.13.一次可同时分析24个元素 (A simultaneous analysis of 24 elements )。