举报
快速选型
OmniPac PCX-500混合模式色谱柱 042195
基本信息:
价格:面议
供货周期:现货
品牌:戴安
货号:

42195

 
参数规格:
具体参数见产品详情
                     
  • 产品介绍
  • 全部规格
  •      
产品介绍

OmniPac PCX-500 Mixed-Mode Column

The OmniPac® PCX-500 column is a macroporous, high-capacity, solvent-compatible column designed to separate cationic and neutral species in a single run. This column combines cation-exchange and polymeric reversed-phase retention mechanisms in a single column.

  • Good for the simultaneous separation of neutral and cationic species

  • Acid and base compatible, pH 0–14

  • 100% solvent compatible

  • Not as effective as the CS12A for inorganic cations

The multimode separation mechanism is achieved by coating a macroporous hydrophobic core for reversed-phase retention with a cation-exchange latex bead layer for cation-exchange retention. The macroporous structure provides an accessible hydrophobic core where reversed-phase retention occurs. The cation-exchange selectivity is provided by a cation-exchange MicroBead layer that coats the hydrophobic core. Since the cation-exchange layer is identical to that of the OmniPac PCX-100, the two columns exhibit similar ion-exchange selectivity. As a result, the OmniPac PXC-100 can be used to develop the ion-exchange separation, and the method can then be transferred to the OmniPac PXC-500 for further refinement to include solvent for the reversed-phase separation. The OmniPac PCX-500 is 100% solvent compatible and should always be run with at least 1% organic solvent in the mobile phase to maintain column life.

The OmniPac PCX-500 may be used for the separation of compounds such as adrenergics, alcohols, anilines, antidepressants, anti-inflammatory agents, plating bath brighteners, cephalosporins, stains and dyes, herbicides, nucleotides and nucleosides, peptides, sulfonamides, and water-soluble vitamins.


OmniPac PCX-500 Mixed-Mode Column  Specifications
DimensionsOmniPac PCX-500 Analytical Column:
4 × 250 mm
OmniPac PCX-500 Guard Column:
4 × 50 mm
Maximum operating pressure4000 psi (27.9 MPa)
Recommended operating temperatureAmbient
Typical operating conditions1.0 mL/min with at least 1% organic solvent
Recommended flow rate: 1.0 mL/min
Mobile phase compatibility100% compatible with common organic solvents. Recommended eluents: 1% solvent. Acid to maintain weak bases in the cationic form, salt (e.g., Na+ for cation exchange).
Resin composition8.5-μm-diameter ethylvinylbenzene/divinylbenzene macroporous substrate (55% cross-linked) agglomerated with 200-nm MicroBead sulfonic acid-functionalized latex (5% cross-linked).
Anion-exchange capacity120 μeq/column (4 × 250 mm analytical column)
Column constructionPEEK with 10-32 threaded ferrule-style end fittings. All components are nonmetallic.
Analytical Columns
OmniPac PCX-500 Analytical Column (4 x 250 mm)042191
Guard Columns
OmniPac PCX-500 Guard Column (4 x 50 mm)042195


OmniPac PCX-500混合模式色谱柱 042195信息由上海希言科学仪器有限公司为您提供,如您想了解更多关于OmniPac PCX-500混合模式色谱柱 042195报价、型号、参数等信息,欢迎来电或留言咨询。

除供应OmniPac PCX-500混合模式色谱柱 042195外,上海希言科学仪器有限公司还可为您提供Wasters 186006220XBridge/XSelect色谱柱Wasters186002051 在线样品前处理柱6025.2012赛默飞UltiMate3000柱塞杆垫 6025.2012 1个/包等产品,公司有专业的客户服务团队,是您值得信赖的合作伙伴。