共有 20 人回复了该问答高压机的灵敏度每年至少需要核查一次
 回复chenboyou发表于:2020/8/1 8:44:02悬赏金额:2积分 状态:未解决
1. 高压机校准 - UL 要求摘录如下(灵敏度要求 - 高压机的灵敏度每年至少需要核查一次)

Sensitivity Requirements - At least once each year the sensitivity of the dielectric equipment is to be checked. The sensitivity of the equipment is to be such that when adjusted to produce the required test voltage and with a 120,000 ohm resistor across the test terminals, the equipment indicates a "breakdown" (a nonconforming performance indication by the tester) within 0.5 seconds. The resistor used shall be provided by the manufacturer and shall have a resistance found to be equal to or greater than 120,000 ohm when measured using a calibrated ohmmeter traceable to the National Institute of Standards and Technology (NIST). If the equipment does not meet this sensitivity test, it shall be readjusted and provision made to maintain the proper calibration.
灵敏度的要求-至少每年一次的介电设备的灵敏度进行检查。该设备的灵敏度是这样的,当调整以产生所要求的试验电压和一个120000欧姆的电阻在测试终端设备,表示“崩溃”(一个不合格的绩效指标的测试)在0.5秒。电阻器必须由制造商提供的,必须有一个电阻的发现是等于或大于120000欧姆欧姆表测量时,使用校准可溯源到国家标准与技术研究所(NIST)。如果设备不符合这种敏感性试验,应当调整,保持适当的校准设置。


该项

《JJG 795-2004 耐电压测试仪检定规程》没作规定,想大家提供更佳方案,先谢!可见http://www.gfjl.org/home/html/08/t-169008.html
 回复  1# chenboyou  回复于:2020-08-02 09:24:02
本人已要求变更《JJG 795-2004 耐电压测试仪检定规程》http://www.gfjl.org/viewthread.php?tid=169008&extra=

希望同行们提出好的建议!